PhotoStability Chamber

 

Technical Specification

  • Temp. Range : 20°C to 60°C
  • Temp. Accuracy : ± 0.2°C
  • Temp. Uniformity : ±  1°C
  • Humidity Range : 40% Rh to 90% Rh
  • Humidity Accuracy : + 2 % Rh
  • Humidity Uniformity : + 3 % Rh
  • Fluorescent tube for day light effect and  UV tubes for Ultra Violet ray in accordance to the ICH guidelines requirement of a lightning energy mounting to 1.2 mil lux per hours and integrated UV energy share of 200 Watt Hours per mtr sqr.
  • As per ICH guidelines :

  • At 25°C : RH = 60%
  • At 30°C : RH = 65%
  • At 40°C : RH = 75%
Standard Model (S)
Inside S.S. 304 with mirror polish & outside mild steel powder coated.
GMP Model (G)
Inside S.S. 316 with mirror polish & outside stainless steel 304 with dull buff finish.

Specification

  • Double walled construction with backside triple.
  • 3” Thick PUF Insulation ensures stable temperature with reduced energy consumption.
  • Motorized Blower at back side of the chamber develops unique air flow system which ensures maximum uniform temperature & humidity condition inside the chamber.U nique air flow assures quick recovery after door openings.
  • 1/12 HP, TEFC , F-class insulation, Single phase, 1440 RPM, 230 Volts
  • High quality U shape S.S. Tubular Heaters are used for better heating.
  • Steam injection system (Non condensing type) & reservoir tank with water level arrangement with indicator.
  • Temperature Control
    • Temperature Control : By means of Imported Microprocessor based auto tuned PID controller with CE mark & dual display of set value & process value.
    • Humidity Control
      • Humidity Control & Sensing : By means of Imported Microprocessor based auto tuned PID controller with CE mark & dual display of set value & process value with direct % RH.Direct RH electronic capacitor type humidity sensor, avoid wick cup & water.
      • Safety Features
        • Additional Safety thermostat to cut of dry heater supply in case temperature of overshoot.
        • Float switch to cutoff the wet heater supply in case of low water level. Audio & Visual alarm.
        • Compressor ON delay timer (2 min.) to safe guard the compressor.
        • In built over load protector provided for hermetically sealed compressor.
      • Refrigeration System
        • Refrigeration by means of CFC Free hermitically sealed Emerson Copland make compressor with R 134A for better cooling with ECO friendly refrigerant with time delay to safe guard compressor system.
      • 50 mm validation port hole with silicon rubber seal to insert sensors for validation.
      • Full view observation Acrylic door with gasket to observe sample inside the chamber.
      • Interior illumination for Working area
      • Specially designed stainless steel rod trays ensure uniform temperature distribution.

      Models
      (S) or (G)

      Capacity
      Liters

      Internal size
      H x W x D cm

      External Size
      H x W x D cm

      Trays

      TI–900 S / G

      100

      50 x 50 x 40

      113 x 71 x 98

      2

      TI–901 S / G

      170

      70 x 60 x 40

      133 x 81 x 98

      2

      TI–902 S /G

      200

      60 x 60 x 60

      123 x 81 x 118

      2

      TI–903 S / G

      285

      80 x 60 x 60

      146 x 81 x 118

      3

      TI–904 S / G

      325

      90 x 60 x 60

      156 x 81 x 118

      3

      TI-905 S / G
      400
      110 x 60 x 60
      176 x 81 x 118
      3

       

      Note   : Two blowers for 400 ltrs to 2000 ltrs. Capacity chambers.

      OPTIONAL FEATURES

      • PLC for auto change over for standby refrigeration system with HMI
      • CFR 21 part 11 compliance software for PLC with RS 485 or Ethernet communication
      • Stand by refrigeration and humidity system
      • Data Logger
      • CFR 21 part 11 compliance software for data logger
      • Data scanner complete with sensor
      • SMS Mobile alert system
      • Door access security system
      • Safety digital temperature controller
      • Timer
      • Lux meter
      • UV meter